Determination of cadmium in solid samples using solid sampling atomic fluorescence spectrometry
-
摘要: 使用由北京吉天仪器有限公司开发的新型固体进样汞镉测定仪(DCMA-200),对采集的土壤、大气颗粒物和固体废物样品中镉的浓度进行分析。在测试过程中,选择9种标准土(GBW07407、GBW07402、GBW07410、GBW07425、GBW07426、GBW07428、GBW07419、GBW07309和GBW07418),研究了进样量对测定的影响,并分别在1.0~10.0 mg称量范围内建立了标准曲线(R 2> 0.93)。当固体样品进样量为10.0 mg时,镉的检出限为0.004 7 mg/kg,测定下限为0.016 mg/kg。利用7种标准土(GBW07418和GBW07419除外)绘制的标准曲线,测得的5种实际土壤中镉的浓度与利用电感耦合等离子体质谱仪(ICP-MS)测得的结果较一致。但与ICP-MS测得的结果相比,利用该固体进样汞镉测定仪测得的大气颗粒物中镉的浓度明显偏高,而危险废物中镉的浓度偏低。因此,建议在检测某地区的固体样品时,尽可能选择该区域和该类固体样品的标准品来设计标准曲线,从而提高测试结果的准确性。Abstract: The concentration of cadmium (Cd) in soils, atmospheric particles, and solid wastes was determined using a new solid sampling atomic fluorescence spectrometry (SSAFS), which was designed by Beijing Titan Instruments Co., Ltd. Nine standard soils(GBW07407, GBW07402, GBW07410, GBW07425, GBW07426, GBW07428, GBW07419, GBW07309 and GBW07418) were chosen to study the influence of sample mass on the measurements, and the standard curves were well fitted (R 2 > 0.93) in the range of 1.0-10.0 mg. When the sample mass was 10.0 mg, the detection limit and determination limit were 0.004 7 and 0.016 mg/kg, respectively. The concentration of Cd in five field samples measured with SSAFS using the seven standard curves (except GBW07418 and GBW07419) was consistent with the value measured with the inductively coupled plasma mass spectrometry (ICP-MS). However, compared with the concentration of Cd determined by ICP-MS, the concentration in atmospheric particles determined by SSAFS was much higher, and the value in solid wastes was smaller. As a result, in order to improve the accuracy of data, it was suggested that the same type of standards or standards in the studied area should be used to establish the standard curves before analyzing solid samples in some area.
-
Key words:
- solid sampling /
- atomic fluorescence spectrometry /
- cadmium (Cd) /
- soil
-
[1] DU P, LIU L, LI F . Distribution and fraction of Cd,Pb,Zn and Cu in the soil of an old smelting area in northeast China[J]. Environmental Geology, 2008,55:205-213. [2] NEZHAD M T K, MOHAMMADI K, GHOLAMI K , et al. Cadmium and mercury in topsoils of Babagorogor watershed,western Iran:distribution,relationship with soil characteristics and multivariate analysis of contamination sources[J]. Geoderma, 2014, 219/220:177-185. [3] WU J, BOYLE E A . Low blank preconcentration technique for the determination of lead,copper,and cadmium in small-volume seawater samples by isotope dilution ICP-MS[J]. Analytical Chemistry, 1997,69(13):2464-2470.
doi: 10.1021/ac961204u pmid: 21639382[4] XU D, FAN W, LÜ H , et al. Simultaneous determination of traces amounts of cadmium,zinc,and cobalt based on UV-vis spectrometry combined with wavelength selection and partial least squares regression[J]. Spectrochimica Acta Part A:Molecular and Biomolecular Spectroscopy, 2014,123:430-435.
doi: 10.1016/j.saa.2013.12.086 pmid: 24412797[5] SOSA V, BARCELÓ C, SERRANO N , et al. Antimony film screen-printed carbon electrode for stripping analysis of Cd(Ⅱ),Pb(Ⅱ),and Cu(Ⅱ) in natural samples[J]. Analytica Chimica Acta, 2015,855:34-40.
doi: 10.1016/j.aca.2014.12.011 pmid: 25542087[6] CHEN J, TEO K C . Determination of cadmium,copper,lead and zinc in water samples by flame atomic absorption spectrometry after cloud point extraction[J]. Analytica Chimica Acta, 2001,450:215-222. [7] JAHROMI E Z, BIDARI A, ASSADI Y , et al. Dispersive liquid-liquid microextraction combined with graphite furnace atomic absorption spectrometry ultra trace determination of cadmium in water samples[J]. Analytica Chimica Acta, 2007,585:305-311.
doi: 10.1016/j.aca.2007.01.007 pmid: 17386679[8] DUARTE A T, DESSUY M B, SILVA M M , et al. Determination of cadmium and lead in plastic material from waste electronic equipment using solid sampling graphite furnace atomic absorption spectrometry[J]. Microchemical Journal, 2010,96(1):102-107. [9] TÖRÖK P, $\dot{Z}$EMBERYOVÁ M . Comparison of chemical modifiers for direct determination of Cd,Cu and Zn in food stuffs by solid-sampling-ETAAS[J]. Food Chemistry, 2012,132(1):554-560.
doi: 10.1016/j.foodchem.2011.10.068 pmid: 26434331[10] OLIVEIRA H R, MESKO M F, VALE M G , et al. Development of methods for the determination of cadmium and thallium in oil shale by-products with graphite furnace atomic absorption spectrometry using direct analysis[J]. Microchemical Journal, 2014,116:55-61. [11] 张晓红, 冯礼, 路东 , 等. 固体进样原子荧光用于测量农产品中的镉[J]. 现代科学仪器, 2012(6):157-161.ZHANG X H, FENG L, LU D , et al. Solid sampling atomic fluorescence spectrometry for cadmium determination in agricultural products[J]. Modern Scientific Instruments, 2012(6):157-161. [12] MAO X, LIU J, HUANG Y , et al. Assessment of homogeneity and minimum sample mass for cadmium analysis in powdered certified reference materials and real rice samples by solid sampling electrothermal vaporization atomic fluorescence spectrometry[J]. Journal of Agricultural and Food Chemistry, 2013,61(4):848-853.
doi: 10.1021/jf3045473[13] 黄亚涛, 毛雪飞, 刘霁欣 , 等. 电热蒸发钨丝在线捕获原子荧光光谱法直接测定菠菜中痕量镉[J]. 分析化学, 2013,41(10):1587-1591.
doi: 10.3724/SP.J.1096.2013.30300HUANG Y T, MAO X F, LIU J X , et al. Direct determination of ultratrace cadmium in ppinach by electrothermal vaporization atomic fluorescence spectrometry using on-line atom trap of tungsten coil[J]. Chinese Journal of Analytical Chemistry, 2013,41(10):1587-1591. doi: 10.3724/SP.J.1096.2013.30300[14] 国家环境保护总局. 生活饮用水标准检验方法金属指标:GB/T 5750.6—2006[S]. 北京: 中国标准出版社, 2006. [15] 卫生部,国家标准化管理委员会. 土壤环境监测技术规范:HJ/T 166—2004[S]. 北京: 中国环境科学出版社, 2004. [16] 环境保护部. 环境监测分析方法标准修订技术导则:HJ 168—2010[S]. 北京: 中国环境科学出版社, 2010.
点击查看大图
计量
- 文章访问数: 1513
- HTML全文浏览量: 82
- PDF下载量: 1258
- 被引次数: 0